Calibration of a Small Volume Ionization Chamber for 60Co Gamma Beams with Low and High Activity Sources

Assef Firnando, Firmansyah and Sri Inang, Sunaryati and Nurman, Rajagukguk and Gatot, Wurdiyanto (2016) Calibration of a Small Volume Ionization Chamber for 60Co Gamma Beams with Low and High Activity Sources. Proceeding of 2nd International Conference on SERIR and 14th SPERA, 2. pp. 106-111. ISSN 978-602-70092-0-2

[img]
Preview
Text
assef.pdf

Download (211kB) | Preview

Abstract

This paper describes the measurement for calibrating a PinPoint 3D chamber type of PTW 31016 serial number 004507 with volume of 0.016 cc in term of absorbed dose to water for 60Co gamma beams using low and high activity sources. Measurement has been carried out using a Substitution Methods at the reference condition. A 0.6 cc ionization chamber type of TW 30013 serial number 6367 was used as a standard chamber connected to Farmer type of NE 2570/1B serial number 1182 and Webline type of T 10022 serial number 268 electrometers with traceability to the Bureau International des poids measures (BIPM) France. Testing on the calibration result were also described based on the determination of the absorbed dose to water at the reference conditions for a 10 MV photon beam and 60Co gamma beams.The results obtained show that the absorbed dose to water calibration factor for the PinPoint3D chamber connected to a Farmer electrometer were ND,w= 2803.2 mGy/nC and ND,w= 2533.2 mGy/nC each for low and high activity sources, while connected to Webline were ND,w = 2534.64 mGy/nC dan ND,w = 2516.0 mGy/nC each for low and high activity sources. The test results for the absorbed dose to water determination for a 10 MV photon beam and Co-60 gamma beams show good agreement by 0.95% and 1.68%. As Summary, the calibration of the PinPoint 3D chamber for 60Co gamma beams using low and high activity sources obtained good results, except for a low activity source, if the PinPoint 3D chamber connected to a Farmer electrometer.

Item Type: Article
Subjects: Keselamatan dan Keamanan Nuklir > Keselamatan Radiasi dan Kesehatan kerja
Divisions: Pusat Teknologi Keselamatan dan Metrologi Radiasi
Depositing User: USER PTKMR BATAN
Date Deposited: 27 Nov 2018 11:56
Last Modified: 27 Nov 2018 11:56
URI: http://repo-nkm.batan.go.id/id/eprint/3926

Actions (login required)

View Item View Item